Show simple item record

dc.contributor.authorHammerschlag, M R
dc.contributor.authorGelling, M
dc.contributor.authorRoblin, P M
dc.contributor.authorWorku, M
dc.date.accessioned2023-06-07T15:45:53Z
dc.date.available2023-06-07T15:45:53Z
dc.date.issued1990-06
dc.identifier.citationHammerschlag MR, Gelling M, Roblin PM, Worku M. Comparison of Kodak Surecell Chlamydia Test Kit with culture for the diagnosis of chlamydial conjunctivitis in infants. J Clin Microbiol. 1990 Jun;28(6):1441-2. doi: 10.1128/jcm.28.6.1441-1442.1990. PMID: 2199506; PMCID: PMC267949.en_US
dc.identifier.issn0095-1137
dc.identifier.pmid2199506
dc.identifier.urihttp://hdl.handle.net/20.500.12648/9909
dc.description.abstractThe accuracy of the Surecell Chlamydia Test Kit (Kodak Clinical Products, Rochester, N.Y.) in detecting neonatal conjunctival infection caused by Chlamydia trachomatis was determined by comparison of this enzyme immunoassay with the isolation of C. trachomatis in tissue culture. Kodak Surecell is a rapid monoclonal antibody-based membrane capture enzyme immunoassay which can be processed in the office of a physician. The sensitivity and specificity compared to culture in detecting C. trachomatis in conjunctival specimens from infants with conjunctivitis were 93 and 96%, respectively. The test does not require specialized equipment or trained personnel and would be ideal for physicians who see low numbers of infants with possible chlamydial conjunctivitis in their offices.
dc.language.isoenen_US
dc.relation.urlhttps://journals.asm.org/doi/10.1128/jcm.28.6.1441-1442.1990en_US
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 International*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.titleComparison of Kodak Surecell Chlamydia Test Kit with culture for the diagnosis of chlamydial conjunctivitis in infants.en_US
dc.typeArticle/Reviewen_US
dc.source.journaltitleJournal of clinical microbiologyen_US
dc.source.volume28
dc.source.issue6
dc.source.beginpage1441
dc.source.endpage2
dc.source.countryUnited States
dc.description.versionVoRen_US
refterms.dateFOA2023-06-07T15:45:54Z
html.description.abstractThe accuracy of the Surecell Chlamydia Test Kit (Kodak Clinical Products, Rochester, N.Y.) in detecting neonatal conjunctival infection caused by Chlamydia trachomatis was determined by comparison of this enzyme immunoassay with the isolation of C. trachomatis in tissue culture. Kodak Surecell is a rapid monoclonal antibody-based membrane capture enzyme immunoassay which can be processed in the office of a physician. The sensitivity and specificity compared to culture in detecting C. trachomatis in conjunctival specimens from infants with conjunctivitis were 93 and 96%, respectively. The test does not require specialized equipment or trained personnel and would be ideal for physicians who see low numbers of infants with possible chlamydial conjunctivitis in their offices.
dc.description.institutionSUNY Downstateen_US
dc.description.departmentPediatricsen_US
dc.description.degreelevelN/Aen_US
dc.identifier.journalJournal of clinical microbiology


Files in this item

Thumbnail
Name:
jcm00054-0376.pdf
Size:
357.5Kb
Format:
PDF

This item appears in the following Collection(s)

Show simple item record

Attribution-NonCommercial-NoDerivatives 4.0 International
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-NoDerivatives 4.0 International